What are the steps for the calibration of the secondary element measuring instrument?

信息来源于:互联网 发布于:2022-04-15

What are the steps for the calibration of the secondary element measuring instrument? The two-dimensional measuring instrument product not only measures the two-dimensional plane size, but also detects the steps and depth of the product. The accuracy of angle detection has always been our consideration. The two-dimensional image measuring instrument can measure the regression line with small deviation.
  1. Open the computer, quadratic element and other related equipment one by one, and click the measurement desktop icon of the quadratic element.

2. Put the calibration version on the test bench with the characters 2014 facing upwards. The calibration version should be placed on the test bench horizontally without foreign objects blocking it or standing on tiptoe.


3. Find the circle of the calibration board, turn off the upper LED light, adjust the resolution of the debugging lens, and adjust the calibration light (lower light).

4. Click the measurement desktop icon "Image Correction - Three Circles under the Correction Processing Menu", then three circles will appear on the screen. Shake the three circles on the screen to the edge or inside of the reference circle in turn, and each time you shake a circle Click the left button of the mouse and a blue circle will appear (Note: the reference circle should not be out of the screen), then click the right mouse button and the screen will display the calibration success and magnification, click OK and then measure the reference circle.

5. After the calibration is completed, measure the round hole on the standard part, click the automatic capture edge point and circle command during measurement, and then click the right mouse button to capture more than three points on the edge of the reference circle to draw a circle, and the size will be automatically generated. If the measured size is consistent with the standard size (tolerance ±0.005mm), the equipment is in good condition; otherwise, if it exceeds the tolerance range, the equipment may be abnormal, and you should contact the two-dimensional imager supplier in time.

6. After each calibration, the calibration personnel will record the magnification, calibration time, personnel and the error value of the calibration, for the measurement personnel to consider the error value when actually testing.